Specifications
PROVA-6830A + 6801 (100A)
General Specifications: Indoors Use
PROVA 6830A Analyzer
| Battery Type: |
1.5V SUM-3 x 8 |
| External DC Input: |
Use only power supply adapter
Model PHAPSA |
| Display: |
Dot Matrix LCD (240×128) with backlight |
| LCD Update Rate: |
1 time / second |
| Power Consumption: |
140mA (approx.) |
| No. Of Samples: |
1024 samples / period |
| Data Logging Files: |
85 |
| Max. File Capacity: |
17474 records (3P4W, 3P3W)
26210 records (1P3W)
52420 records (1P2W)
4096 records (50 Harmonics / record) |
| Sampling Time: |
2 to 3000 seconds for data logging |
| Low battery Indication: |
|
| Overload Indication: |
OL |
| Operating Temperature: |
-10°C to 50°C |
| Operating Humidity: |
less than 85% relative |
| Storage Temperature: |
-20°C to 60°C |
| Storage Humidity: |
less than 75% relative |
| Dimension: |
257 (L) x 155 (W) x 57 (H) mm
10.1” (L) x 6.1” (W) x 2.3” (H) |
| Weight: |
1160g (Batteries included) |
| Accessories: |
Probes (model 6801) x 3
Test leads (3 meter long) x 4
Alligator clips x 4
Carrying bag x 1
Users manual x 1
Batteries 1.5V x 8
AC power adapter x 1 Software CD x 1 Software users manual x 1 USB to RS232 cable x 1 |
Electrical Specifications: (23℃±5℃)
AC Watt
| Range (0 to 100A) |
Resolution |
Accuracy of Readings |
| 5.0 – 999.9 W |
0.1W |
±1% ± 0.8W |
| 1.000 – 9.999 KW |
0.001 KW |
±1% ± 8W |
| 10.00 – 99.99 KW |
0.01 KW |
±1% ± 80W |
| 100.0 – 999.9 KW |
0.1 KW |
±1% ± 0.8KW |
| 1000 – 9999 KW |
1 KW |
±1% ± 8KW |
Range of CT (Current Transformer) Ratio: 1 to 600
AC Apparent Power (VA, from 0.000 VA to 9999 KVA): VA = V r.m.s. x A r.m.s
AC Reactive Power (VAR, from 0.000 VAR to 9999 KVAR): VAR = √(VA2 – W2)
AC Active Energy (mWH, WH, or KWH, from 0mWH to 999,999KWH): WH = W * Time (in hours)
AC Current (Auto range, TRMS, Overload Protection AC 200A)
| Range |
Resolution |
Accuracy of Readings |
| 0.04 – 1 A |
0.1mA / 1mA |
±0.5% ± 0.05A |
| 0.4 – 10 A |
0.001A / 0.01A |
±0.5% ± 0.05A |
| 4 – 100 A |
0.01A / 0.1A |
±1.0% ± 0.5A |
AC Voltage (Auto range, TRMS, Overload Protection AC 800V)
| Range |
Resolution |
Accuracy of Readings |
| 20.0 V – 500.0 V (Phase to Neutral) |
0.1 V |
±0.5% ± 5dgts |
| 20.0 V – 600.0 V (Phase to Phase) |
Range of VT (Voltage Transformer) Ratio: 1 to 3000
Harmonics of AC Voltage in Percentage
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% |
| 21 – 49th |
±4% of reading ± 2.0% |
| 50 – 99th |
±6% of reading ± 2.0% |
Harmonics of AC Voltage in Magnitude
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1V |
±2% ± 0.5V |
| 21 – 49th |
±4% of reading ± 0.5V |
| 50 – 99th |
±6% of reading ± 0.5V |
Harmonics of AC Current in Percentage
| Range |
Resolution |
Accuracy |
| 1 – 10th |
0.1% |
±0.2% of reading ± 1% |
| 11 – 20th |
±2% of reading ± 1% |
| 21 – 50th (A range) |
±5% of reading ± 1% |
| 21 – 50th (mA range) |
±10% of reading ± 1% |
| 51 – 99th |
±35% of reading ± 1% |
Harmonics of AC Current in Magnitude
| Range |
Resolution |
Accuracy |
| 1 – 10th |
0.1mA / 0.1A |
±0.2% of reading ±7dgts |
| 11 – 20th |
±2% of reading ±7dgts |
| 21 – 50th (A range) |
±5% of reading ±7dgts |
| 21 – 50th (mA range) |
±10% of reading ±7dgts |
| 51 – 99th |
±35% of reading ±7dgts |
Power Factor (PF)
| Range |
Resolution |
Accuracy |
| 0.00 – 1.00 |
0.01 |
± 0.04 |
Phase Angle (Φ)
| Range |
Resolution |
Accuracy |
| -180° to 180° |
0.1° |
± 1° |
Peak Value of ACV (peak value > 20V) or ACA (peak value > 10% of the range), VT=1
| Range |
Sampling Time |
Accuracy of Reading |
| 50 Hz |
19μs |
± 5% ± 30 digits |
| 60 Hz |
16μs |
Crest Factor (C.F.) of ACV (peak value >20V)
or ACA (peak value > 10% of the range), VT=1
| Range |
Resolution |
Accuracy of Readings |
| 1.00 – 99.99 |
0.01 |
± 5% ± 30 digits |
Frequency in AUTO mode
| Range |
Resolution |
Accuracy |
| 45 – 65 Hz |
0.1Hz |
± 0.1Hz |
Total Harmonic Distortion (THD-F)
| Range |
Resolution |
Accuracy |
| 0.0 – 20.0 % |
0.1% |
± 1% |
| 20.0 – 100% |
±3% of reading ± 5% |
| 100 – 999.9% |
±10% of reading ±10% |
PROVA 6801 Current Probes
| Conductor Size: |
30mm (approx.) |
| Cable Length: |
3000mm |
| Range Selection: |
Manual (1A, 10A, 100A) |
| Battery: |
powered by power analyzer |
| Dimension: |
210mm (L) x 62mm (W) x 36mm (H)
8.3″ (L) x 2.5″ (W) x 1.4″ (H) |
| Weight: |
200g |
PROVA-6830A + 6802 (1000A)
General Specifications: Indoors Use
PROVA 6830A Analyzer
| Battery Type: |
1.5V SUM-3 x 8 |
| External DC Input: |
Use only power supply adapter
Model PHAPSA |
| Display: |
Dot Matrix LCD (240×128) with backlight |
| LCD Update Rate: |
1 time / second |
| Power Consumption: |
140mA (approx.) |
| No. Of Samples: |
1024 samples / period |
| Data Logging Files: |
85 |
| Max. File Capacity: |
17474 records (3P4W, 3P3W)
26210 records (1P3W)
52420 records (1P2W)
4096 records (50 Harmonics / record) |
| Sampling Time: |
2 to 3000 seconds for data logging |
| Low battery Indication: |
|
| Overload Indication: |
OL |
| Operating Temperature: |
-10°C to 50°C |
| Operating Humidity: |
less than 85% relative |
| Storage Temperature: |
-20°C to 60°C |
| Storage Humidity: |
less than 75% relative |
| Dimension: |
257 (L) x 155 (W) x 57 (H) mm
10.1” (L) x 6.1” (W) x 2.3” (H) |
| Weight: |
1160g (Batteries included) |
| Accessories: |
Probes (model 6801) x 3
Test leads (3 meter long) x 4
Alligator clips x 4
Carrying bag x 1
Users manual x 1
Batteries 1.5V x 8
AC power adapter x 1 Software CD x 1 Software users manual x 1 USB to RS232 cable x 1 |
Electrical Specifications: (23℃±5℃)
AC Watt
| Range (0 to 1000A) |
Resolution |
Accuracy of Readings |
| 5.0 – 999.9 W |
0.1W |
±1% ± 0.8W |
| 1.000 – 9.999 KW |
0.001 KW |
±1% ± 8W |
| 10.00 – 99.99 KW |
0.01 KW |
±1% ± 80W |
| 100.0 – 999.9 KW |
0.1 KW |
±1% ± 0.8KW |
| 1000 – 9999 KW |
1 KW |
±1% ± 8KW |
| 0.000 – 9.999MW |
0.001MW |
±1% ± 80KW |
Range of CT (Current Transformer) Ratio: 1 to 600
AC Apparent Power (VA, from 0.000 VA to 9999 KVA): VA = V r.m.s. x A r.m.s
AC Reactive Power (VAR, from 0.000 VAR to 9999 KVAR): VAR = √(VA2 – W2)
AC Active Energy (mWH, WH, or KWH, from 0mWH to 999,999KWH):
WH = W * Time (in hours)
AC Current (Auto range, TRMS, Overload Protection AC 2000A)
| Range |
Resolution |
Accuracy of Readings |
| 10.00A |
0.001A/0.01A |
– |
| 5A – 100.0A |
0.01A/0.1A |
±0.5% ± 0.5A |
| 50A – 1000.0 A |
0.1A/1 A |
±0.5% ± 5A |
AC Voltage (Auto range, TRMS, Overload Protection AC 800V)
| Range |
Resolution |
Accuracy of Readings |
| 20.0 V – 500.0 V (Phase to Neutral) |
0.1 V |
±0.5% ± 5dgts |
| 20.0 V – 600.0 V (Phase to Phase) |
Range of VT (Voltage Transformer) Ratio: 1 to 3000
Harmonics of AC Voltage in Percentage
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% |
| 21 – 49th |
±4% of reading ± 2.0% |
| 50 – 99th |
±6% of reading ± 2.0% |
Harmonics of AC Voltage in Magnitude
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1V |
±2% ± 0.5V |
| 21 – 49th |
±4% of reading ± 0.5V |
| 50 – 99th |
±6% of reading ± 0.5V |
Harmonics of AC Current in Percentage
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1 % |
±2% |
| 21 – 49th |
±4% of reading ± 2.0% |
| 50 – 99th |
±6% of reading ± 2.0% |
Harmonics of AC Current in Magnitude
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1A |
±2% of reading ±0.4A |
| 21 – 49th |
±4% of reading ±0.4A |
| 50 – 99th |
±6% of reading ±0.4A |
Power Factor (PF)
| Range |
Resolution |
Accuracy |
| 0.00 – 1.00 |
0.01 |
± 0.04 |
Phase Angle (Φ)
| Range |
Resolution |
Accuracy |
| -180° to 180° |
0.1° |
± 1° |
Peak Value of ACV (peak value > 20V) or ACA (peak value > 20A), VT=1
| Range |
Sampling Time |
Accuracy of Reading |
| 50 Hz |
19μs |
± 5% ± 30 digits |
| 60 Hz |
16μs |
Crest Factor (C.F.) of ACV (peak value >20V) or ACA (peak value > 20A), VT=1
| Range |
Resolution |
Accuracy of Readings |
| 1.00 – 99.99 |
0.01 |
± 5% ± 30 digits |
Frequency in AUTO mode
| Range |
Resolution |
Accuracy |
| 45 – 65 Hz |
0.1Hz |
± 0.1Hz |
Total Harmonic Distortion (THD-F)
| Range |
Resolution |
Accuracy |
| 0.0 – 20% |
0.1% |
± 2% |
| 20 – 100% |
± 6% of reading ± 1% |
| 100 – 999.9 % |
± 10% of reading ± 1% |
PROVA 6802 Current Probes
| Conductor Size: |
55mm (approx.), 64 x 24mm (bus bar) |
| Cable Length: |
3000mm |
| Range Selection: |
Manual (10A, 100A, 1000A) |
| Battery: |
powered by power analyzer |
| Dimension: |
244mm (L) x 97mm (W) x 46mm (H)
9.6″ (L) x 3.8″ (W) x 1.8″ (H) |
| Weight: |
600g |
PROVA-6830A + 3007 (3000A)
Electrical Specifications: (23℃±5℃)
AC Watt
| Range
(0 to 3000A) |
Resolution |
Accuracy of Readings |
| > 20 V and > 30A |
< 20V or < 30A |
| 10.0 – 999.9 W |
0.1W |
±1% of range |
±2% of range |
| 1.000 – 9.999 KW |
0.001 KW |
| 10.00 – 99.99 KW |
0.01 KW |
| 100.0 – 999.9 KW |
0.1 KW |
| 1000 – 9999 KW |
1 KW |
Range of CT (Current Transformer) Ratio: 1 to 600
AC Apparent Power (VA, from 0.000 VA to 9999 KVA): VA = V r.m.s. x A r.m.s
AC Reactive Power (VAR, from 0.000 VAR to 9999 KVAR): VAR = √(VA2 – W2)
AC Active Energy (mWH, WH, or KWH, from 0mWH to 999,999KWH): WH = W * Time (in hours)
AC Current (Auto range, TRMS, Overload Protection AC 3000A)
| Range |
Resolution |
Accuracy of Readings |
| 10.0 – 300.0A |
0.01A / 0.1A |
±1% of range |
| 300.0 – 3000A |
0.1A / 1A |
AC Voltage (Auto range, TRMS, Overload Protection AC 800V)
| Range |
Resolution |
Accuracy of Readings |
| 20.0 V – 500.0 V (Phase to Neutral) |
0.1 V |
±0.5% ± 5dgts |
| 20.0 V – 600.0 V (Phase to Phase) |
Range of VT (Voltage Transformer) Ratio: 1 to 3000
Harmonics of AC Voltage in Percentage
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% |
| 21 – 49th |
±4% of reading ± 2.0% |
| 50 – 99th |
±6% of reading ± 2.0% |
Harmonics of AC Voltage in Magnitude
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1V |
±2% ± 0.5V |
| 21 – 49th |
±4% of reading ± 0.5V |
| 50 – 99th |
±6% of reading ± 0.5V |
Peak Value of ACV (peak value > 20V) or ACA (peak value > 30A), VT=1
| Range |
Sampling Time |
Accuracy of Reading |
| 50 Hz |
19μs |
± 5% ± 30 digits |
| 60 Hz |
16μs |
Crest Factor (C.F.) of ACV (peak value >20V) or ACA (peak value > 30A), VT=1
| Range |
Resolution |
Accuracy of Readings |
| 1.00 – 99.99 |
0.01 |
± 5% ± 30 digits |
Harmonics of AC Current in Percentage
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% |
| 21 – 50th |
±6% |
| 51 – 99th |
±10% |
Harmonics of AC Current in Magnitude
(1 to 99th order, min. current at the 50 or 60 Hz, True RMS < 300A)
| Range (0 – 300A) |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% of reading ± 4A |
| 21 – 50th |
±4% of reading ± 4A |
| 51 – 99th |
±6% of reading ± 4A |
(1 to 99th order, min. current at the 50 or 60 Hz, 3000A > True RMS > 300A)
| Range (300–3000A) |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% of reading ± 40A |
| 21 – 50th |
±4% of reading ± 40A |
| 51 – 99th |
±6% of reading ± 40A |
Power Factor (PF)
| Range |
Resolution |
Accuracy |
| > 20V and > 30A |
< 20V or < 30A |
| 0.000 – 1.000 |
0.001 |
± 0.04 |
±0.1 |
Phase Angle (Φ, V > 20V, A > 30A)
| Range |
Resolution |
Accuracy |
| -180° to 180° |
0.1° |
± 2° |
| 0° to 360° |
Frequency of ACV (RMS value > 10V) or ACA (RMS value > 30A)
| Range |
Resolution |
Accuracy |
| 45 – 65 Hz |
0.1 Hz |
± 0.2Hz |
Total Harmonic Distortion (THD-F)
| Range |
Resolution |
Accuracy |
| 0.0 – 20% |
0.1% |
± 2% |
| 20 – 100% |
± 6% of reading ± 5% |
| 100 – 999.9 % |
± 10% of reading ± 10% |
General Specifications: Indoors Use
AFLEX 3007 Flexible Current Probes
| Probe Length: ( 3007-24) |
24 in / 610 mm |
| Minimum Bending Diameter: |
35mm |
| Connector Diameter: |
23mm |
| Cable Diameter: |
14mm |
| Cable Length from Probe to Box: |
1700mm |
| Cable Length from Box to Output: |
1700mm |
| Range Selection: |
Manual (300A, 3000A) |
| Battery: |
powered by power analyzer |
| Dimension (Box): |
130mm(L) x 80mm(W)x 43mm(H)
5.1”(L) x 3.1”(W) x 1.7”(H) |
| Weight: |
410g |
PROVA-6830A + 3009 (1200A)
Electrical Specifications: (23℃±5℃)
AC Watt
| Range
(0 to 1200A) |
Resolution |
Accuracy of Readings |
| > 20 V and > 30A |
< 20V or < 30A |
| 10.0 – 999.9 W |
0.1W |
±1% of range |
±2% of range |
| 1.000 – 9.999 KW |
0.001 KW |
| 10.00 – 99.99 KW |
0.01 KW |
| 100.0 – 999.9 KW |
0.1 KW |
| 1000 – 9999 KW |
1 KW |
Range of CT (Current Transformer) Ratio: 1 to 600
AC Apparent Power (VA, from 0.000 VA to 9999 KVA): VA = V r.m.s. x A r.m.s
AC Reactive Power (VAR, from 0.000 VAR to 9999 KVAR): VAR = √(VA2 – W2)
AC Active Energy (mWH, WH, or KWH, from 0mWH to 999,999KWH): WH = W * Time (in hours)
AC Current (Auto range, TRMS, Overload Protection AC 1200A)
| Range |
Resolution |
Accuracy of Readings |
| 6.0 – 120.0A |
0.01A / 0.1A |
±1% of range |
| 120.0 – 1200A |
0.1A / 1A |
AC Voltage (Auto range, TRMS, Overload Protection AC 800V)
| Range |
Resolution |
Accuracy of Readings |
| 20.0 V – 500.0 V (Phase to Neutral) |
0.1 V |
±0.5% ± 5dgts |
| 20.0 V – 600.0 V (Phase to Phase) |
Range of VT (Voltage Transformer) Ratio: 1 to 3000
Harmonics of AC Voltage in Percentage
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% |
| 21 – 49th |
±4% of reading ± 2.0% |
| 50 – 99th |
±6% of reading ± 2.0% |
Harmonics of AC Voltage in Magnitude
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1V |
±2% ± 0.5V |
| 21 – 49th |
±4% of reading ± 0.5V |
| 50 – 99th |
±6% of reading ± 0.5V |
Peak Value of ACV (peak value > 20V) or ACA (peak value > 30A), VT=1
| Range |
Sampling Time |
Accuracy of Reading |
| 50 Hz |
19μs |
± 5% ± 30 digits |
| 60 Hz |
16μs |
Crest Factor (C.F.) of ACV (peak value >20V) or ACA (peak value > 30A), VT=1
| Range |
Resolution |
Accuracy of Readings |
| 1.00 – 99.99 |
0.01 |
± 5% ± 30 digits |
Harmonics of AC Current in Percentage
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% |
| 21 – 50th |
±6% |
| 51 – 99th |
±10% |
Harmonics of AC Current in Magnitude
(1 to 99th order, min. current at the 50 or 60 Hz, True RMS < 120A)
| Range (0 – 120A) |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% of reading ± 4A |
| 21 – 50th |
±4% of reading ± 4A |
| 51 – 99th |
±6% of reading ± 4A |
(1 to 99th order, min. current at the 50 or 60 Hz, 1200A > True RMS > 120A)
| Range (120–1200A) |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% of reading ± 40A |
| 21 – 50th |
±4% of reading ± 40A |
| 51 – 99th |
±6% of reading ± 40A |
Power Factor (PF)
| Range |
Resolution |
Accuracy |
| > 20V and > 30A |
< 20V or < 30A |
| 0.000 – 1.000 |
0.001 |
± 0.04 |
±0.1 |
Phase Angle (Φ, V > 20V, A > 30A)
| Range |
Resolution |
Accuracy |
| -180° to 180° |
0.1° |
± 2° |
| 0° to 360° |
Frequency of ACV (RMS value > 10V) or ACA (RMS value > 30A)
| Range |
Resolution |
Accuracy |
| 45 – 65 Hz |
0.1 Hz |
± 0.2Hz |
Total Harmonic Distortion (THD-F)
| Range |
Resolution |
Accuracy |
| 0.0 – 20% |
0.1% |
± 2% |
| 20 – 100% |
± 6% of reading ± 5% |
| 100 – 999.9 % |
± 10% of reading ± 10% |
AFLEX 3009 Flexible Current Probes
| Probe Length: ( 3009-18) |
18 in / 460 mm |
| Minimum Bending Diameter: |
35mm |
| Connector Diameter: |
23mm |
| Cable Diameter: |
14mm |
| Cable Length from Probe to Box: |
1700mm |
| Cable Length from Box to Output: |
1700mm |
| Range Selection: |
Manual (120A, 1200A) |
| Battery: |
powered by power analyzer |
| Dimension (Box): |
130mm(L) x 80mm(W)x 43mm(H)
5.1”(L) x 3.1”(W) x 1.7”(H) |
| Weight: |
390g |
PROVA-6830A + 3006 (6000A)
Electrical Specifications: (23℃±5℃)
AC Watt
| Range(0 to 6000A) |
Resolution |
Accuracy of Readings |
| > 20 V and > 60A |
< 20V or < 60A |
| 10.0 – 999.9 W |
0.2W |
±1% of range |
±2% of range |
| 1.000 – 9.999 KW |
0.002 KW |
| 10.00 – 99.99 KW |
0.02 KW |
| 100.0 – 999.9 KW |
0.2 KW |
| 1000 – 9999 KW |
2 KW |
Range of VT (Voltage Transformer) Ratio: 1 to 3000
AC Apparent Power (VA, from 0.000 VA to 9999 KVA): VA = V r.m.s. x A r.m.s
AC Reactive Power (VAR, from 0.000 VAR to 9999 KVAR): VAR = √(VA2 – W2)
AC Active Energy (mWH, WH, or KWH, from 0mWH to 999,999KWH): WH = W * Time (in hours)
AC Current (Auto range, TRMS, Overload Protection AC 6000A)
| Range |
Resolution |
Accuracy of Readings |
| 20.0 – 600.0A |
0.2A |
±1% of range |
| 600.0 – 6000A |
0.2A / 2A |
AC Voltage (Auto range, TRMS, Overload Protection AC 800V)
| Range |
Resolution |
Accuracy of Readings |
| 20.0 V – 500.0 V (Phase to Neutral) |
0.1 V |
±0.5% ± 5dgts |
| 20.0 V – 600.0 V (Phase to Phase) |
Range of VT (Voltage Transformer) Ratio: 1 to 3000
Harmonics of AC Voltage in Percentage
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% |
| 21 – 49th |
±4% of reading ± 2.0% |
| 50 – 99th |
±6% of reading ± 2.0% |
Harmonics of AC Voltage in Magnitude
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1V |
±2% ± 0.5V |
| 21 – 49th |
±4% of reading ± 0.5V |
| 50 – 99th |
±6% of reading ± 0.5V |
Peak Value of ACV (peak value > 20V) or ACA (peak value > 60A), VT=1
| Range |
Sampling Time |
Accuracy of Reading |
| 50 Hz |
19μs |
± 5% ± 30 digits |
| 60 Hz |
16μs |
Crest Factor (C.F.) of ACV (peak value >20V) or ACA (RMS value > 60A), VT=1
| Range |
Resolution |
Accuracy of Readings |
| 1.00 – 99.99 |
0.01 |
± 5% ± 60 digits |
Harmonics of AC Current in Percentage
| Range |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% |
| 21 – 50th |
±6% |
| 51 – 99th |
±10% |
Harmonics of AC Current in Magnitude
(1 to 99th order, min. current at the 50 or 60 Hz, True RMS < 600A)
| Range (0 – 600A) |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% of reading ± 8A |
| 21 – 50th |
±4% of reading ± 8A |
| 51 – 99th |
±6% of reading ± 8A |
(1 to 99th order, min. current at the 50 or 60 Hz, 6000A > True RMS > 600A)
| Range (600–6000A) |
Resolution |
Accuracy |
| 1 – 20th |
0.1% |
±2% of reading ± 80A |
| 21 – 50th |
±4% of reading ± 80A |
| 51 – 99th |
±6% of reading ± 80A |
Power Factor (PF)
| Range |
Resolution |
Accuracy |
| > 20V and > 60A |
< 20V or < 60A |
| 0.000 – 1.000 |
0.001 |
± 0.04 |
±0.1 |
Phase Angle (Φ, V > 20V, A > 60A)
| Range |
Resolution |
Accuracy |
| -180° to 180° |
0.1° |
± 2° |
| 0° to 360° |
Frequency of ACV (RMS value > 10V) or ACA (RMS value > 60A)
| Range |
Resolution |
Accuracy |
| 45 – 65 Hz |
0.1 Hz |
± 0.2Hz |
Total Harmonic Distortion (THD-F)
| Range |
Resolution |
Accuracy |
| 0.0 – 20% |
0.1% |
± 2% |
| 20 – 100% |
± 6% of reading ± 5% |
| 100 – 999.9 % |
± 10% of reading ± 10% |
General Specifications: Indoors Use
AFLEX 3006 Flexible Current Probes
| Probe Length: ( 3006-36 ) |
36 in / 920 mm |
| Minimum Bending Diameter: |
35mm |
| Connector Diameter: |
23mm |
| Cable Diameter: |
14mm |
| Cable Length from Probe to Box: |
1700mm |
| Cable Length from Box to Output: |
1700mm |
| Range Selection: |
Manual (600A, 6000A) |
| Battery: |
powered by power analyzer |
| Dimension (Box): |
130mm(L) x 80mm(W)x 43mm(H)
5.1”(L) x 3.1”(W) x 1.7”(H) |
| Weight: |
430g |