Phone

+603-5621 5786

Business Hour

08:30 - 17:30

TES DC/AC Clamp Meter PROVA-A1/PROVA-A3

PROVA-A1/PROVA-A3
DC/AC Clamp Meter
Product Features
  • Five kinds of Dual display:
    Hold + Average
    Max + Min
    ΔMax + ΔMin
    Peak Max + Peak Min
    Δ Peak Max + Δ Peak Min
  • With Backlight
  • Auto Range: V, A, Resistance, Capacitance and Temperature.
  • Auto-Power-off function (30 min)
  • Ideal for works in crowded switch box or cable areas

Description

Electrical Specifications: (23℃±5℃)

DC Current

Range Resolution Accuracy Overload Protection
60 A 10mA ±1.5%±2dgts
DC 400A
0-150 A 100mA ±1.5%±2dgts
150-300 A 100mA ±2.2%±2dgts
300-400 A 100mA ±4%±2dgts

DCA Peak MAX ±400A   Accuracy + (±2%±50dgts)

AC Current  (A1: True RMS, A3: Average Sensing)

Range Resolution Accuracy Overload Protection
50/60 Hz 40 – 500Hz
60A 10mA ±1.5%±3dgts ±2.0%±4dgts AC 400A
0-150A 100mA ±1.5%±3dgts ±2.0%±4dgts
150-300A 100mA ±2.2%±3dgts ±2.5%±4dgts
300-400A 100mA 4%±3dgts 5%±4dgts

ACA Peak MAX ±400A   Accuracy + (2%±50dgts)

LPF @1KHz -3db

DC Voltage  (Overload Protection 600V)

Range Resolution Accuracy Input Impedance
60V 10 mV ±1.5%±3dgts 10MΩ
600V 100 mV

DCV Peak MAX ±600V   Accuracy + (±2%±50dgts)

AC Voltage  (A1: True RMS, A3: Average Sensing) (Input Impedance 10 MΩ)

Range Resolution Accuracy Overload Protection
50/60 Hz 40 – 1KHz
60V 10mV ±1.5%±5dgts ±2.0%±5dgts AC 600V
600V 100mV

ACV Peak MAX ±850V   Accuracy + (±2%±50dgts)

LPF @1KHz -3db 

Continuity  (Open voltage 0.4V, Overload protection AC 600V)

Range Resolution Accuracy Beeping
60-600Ω 0.1Ω ±1.0%±2dgts <30.0Ω (approx.)

 

Resistance (Ω (Open voltage 0.4V)

Range Resolution Accuracy Overload Protection
600Ω 0.1Ω ±1.5%±2dgts AC 600V
6KΩ 1Ω
60KΩ 10Ω
600KΩ 100Ω

 

Frequency (auto range)

Range (Hz) Resolution (Hz) Accuracy Sensi-tivity OverloadProtection
4-99.99 0.01 ±0.5%±0.02 V:3V/10V  A:1A/5A AC 600V

 AC400A

100.0~999.9 0.1 ±0.5%±0.2 V:2V/10V  A:1A/5A
1.000K-9.999K 10 ±0.5%±20 V:2V/10V  A:1A/5A
10.00K-99.99K 1.2K ±0.5%±2.4K V:2V/10V  A:1A/5A
100.0K-200.0K 5K ±0.5%±10K V:2V/10V  A:1A/5A

 Temperature  (for using K-Type thermocouples, °C or °F )

Range Resolution Accuracy Overload Protection
-40 – 0 °C 0.1 °C ±2.0%±2 °C AC 300V
0 – 400 °C 0.1 °C ±0.5%±2 °C
-40 – 32 °F 0.1 °F ±2%±3.6 °F
32 – 752 °F 0.1 °F ±0.5%±3.6°F

 Diode Test

Range Resolution Accuracy Overload Protection
0 – 1.999V 0.001V ±2.5%±5dgts AC 600V

 Capacitance

Range Resolution Accuracy Overload Protection
600nF 0.1nF ±2.0%±0.2 nF  

 

AC 600V

F 1nF ±2.0%±0.02μF
60μF 10nF ±2.0%±0.02μF
600μF 100nF ±2.0%±0.2μF
6000μF F ±2.5%±5μF

 

General Specifications:     Indoor Use

Conductor Size:                25mm max. (approx.)

Battery Type:                    two 1.5V SUM-3

Display:                             6000 counts

Range Selection:              Auto and Manual

Overload Indication:         OL

Power Consumption:        17 mA (approx.)

Low battery Indication:     

Sampling Time:                2 times/sec. 

Auto-power-off:                30 minutes

Operating Temperature:  0°C to 50°C

Operating Humidity:        less than 85% relative

Altitude:                           up to 2000M

Storage Temperature:    -20°C to 60°C

Storage Humidity:          less than 75% relative

Dimension:                    190mm (L) x 66mm (W) x 36mm (H)

                                       7.5″ (L) x 2.6″ (W) x 1.45″ (H)

Weight:                           250g (battery included)

Accessories:                  Carrying bag x 1

                                      Users manual x 1

                                      1.5V battery x 2

                                      Test leads x 1 
                                       K-type thermocouples x 1                              

                                       Adapter (for Capacitance and K-type thermocouples) x 1

Ready to Transform Your Business? Contact Us Today

Categories

Request a Quote or Further Information

Please fill in the request form below and one of our engineers will contact you to discuss your system requirements and if we can help, our typical response time is within 1 – 3 working day.