TES AFLEX-6300 Graphic Power Quality Analyzer

The AC Power & Harmonics Analyzer accurately measures AC current up to 3000 A and AC voltage up to 600 V, performs power quality analysis, and logs up to 512K of transient events for industrial and laboratory applications.

  • AC current measurement: 3–3000 A True RMS, 1% basic accuracy
  • AC voltage measurement: 4–600 V True RMS, 0.5% basic accuracy
  • Power measurement: Active (W, KW, HP), Reactive (VAR, KVAR), Apparent (VA, KVA)
  • Peak detection & fast sampling: 33–39 μs

Description

The Advanced AC Power & Harmonics Analyzer delivers precise AC current and voltage measurements. It reads currents from 3 A to 3000 A and voltages from 4 V to 600 V, providing reliable results for active, reactive, and apparent power. Users can also measure phase angle (Φ) and power factor (PF) for both single-phase and balanced three-phase systems.

This analyzer offers harmonic analysis up to the 50th order and tracks total harmonic distortion (%THD). It captures peak voltage and current events with fast sampling of 33–39 μs. The device logs 32,000 transient events in its 512 KB memory, allowing engineers to monitor swell, dip, and outage events. Its 128 × 64 dot matrix LCD with backlight ensures easy reading of data. Users can connect it to a PC via USB-RS232 for logging and analysis. Lightweight and portable, it is ideal for field testing, maintenance, and calibration tasks.

Features

  • Probe & Cable: 610 mm probe, minimum bending diameter 35 mm, cable length 1.7 m

  • Display: 128 × 64 dot matrix LCD with backlight, update rate 2/sec

  • Battery: 2 × 1.5 V SUM-3, low-power consumption (~10 mA)

  • Auto Range & Overload Indication: Automatic range selection with OL alert

  • Data Logging: 512 KB memory, 32,000 maximum transient events

  • Operating Environment: -10 °C to 85 °C, 15–85% RH, altitude up to 2000 m

  • AC Watt/VA/VAR Measurement: Up to 9999 KW/KVA/KVAR, ±1% accuracy

  • Power Factor (PF) & Phase Angle (Φ): Accurate measurements for system analysis

  • Harmonics & THD: 1st–50th order, 0.1% resolution

  • Peak & Fast Sampling: 33 μs (60 Hz), 39 μs (50 Hz)

  • Accessories Included: Test leads, carrying case, batteries, software CD & manual, USB-RS232 cable

  • Auto Power-Off: 15 minutes to preserve battery

Specifications

 

General Specifications:

Probe Length: 6300-24 24 in / 610 mm
Minimum Bending Diameter: 35mm
Connector Diameter: 23mm
Cable Diameter: 14mm
Cable Length from Probe to Box: 1700mm
Cable Length from Box to Output: 1700mm

 

Data Logging Memory Size: 512K bytes
Maximum Transient Events: 32000
Dimension (Box): 130 (L) x 80 (W) x 43 (H) mm

5.1” (L) x 3.1” (W) x 1.7” (H)

Battery Type: two 1.5V SUM-3
Display: 128 X 64 Dot Matrix LCD with Backlight
Range Selection: Auto
Overload Indication: OL
Power Consumption: 10mA (approx)
Auto-Power-Off: 15 minutes after power-on
LCD Update Time: 2 times/sec.
No. Of Samples per Period 512 (voltage or current); 256 (power)
Operating Temperature: -10°C to 85°C
Operating Humidity: 15% to 85% RH
Altitude: up to 2000M
Storage Temperature: -20°C to 85°C
Storage Humidity: 15% to 85% RH
Weight: 430g (battery included)
Accessories: Test leads (red/black) x 1 set Users manual x 1

Batteries 1.5V x 2 Carrying Case x 1 Software CD x 1

Software users manual x 1

USB to RS232 cable x 1

 

Electrical Specifications: (23±5)

AC Watt

 

Range Resolution Accuracy of Readings
> 20 V and > 30A < 20V or < 30A
10.0 – 999.9 W 0.1W ±1% of Range ±0.3W ±2% of Range
1.000 – 9.999 KW 0.001 KW ±1% of Range ±0.003KW ±2% of Range
10.00 – 99.99 KW 0.01 KW ±1% of Range ±0.03KW ±2% of Range
100.0 – 999.9 KW 0.1 KW ±1% of Range ±0.3KW ±2% of Range
1000 – 9999 KW 1 KW ±1% of Range ±3KW ±2% of Range

AC Apparent Power (VA)

 

Range Resolution Accuracy of Readings
> 20 V and > 30A <20V or < 30A
10.0 – 999.9 VA 0.1VA ±1% of Range ±0.3VA ±2% of Range
1.000 – 9.999 KVA 0.001 KVA ±1% of Range ±0.003KVA ±2% of Range
10.00 – 99.99 KVA 0.01 KVA ±1% of Range ±0.03KVA ±2% of Range
100.0 – 999.9 KVA 0.1 KVA ±1% of Range ±0.3KVA ±2% of Range
1000 – 9999 KVA 1 KVA ±1% of Range ±3KVA ±2% of Range

 

AC Reactive Power (VAR)

 

 

Range Resolution Accuracy of Readings
> 20 V and > 30A <20V or < 30A

 

10.0–999.9 VAR 0.1VAR ±1% of Range ±0.3VAR ±2% of Range
1.000–9.999 KVAR 0.001 KVAR ±1% of Range ±0.003KVAR ±2% of Range
10.00–99.99 KVAR 0.01 KVAR ±1% of Range ±0.03KVAR ±2% of Range
100.0–999.9 KVAR 0.1 KVAR ±1% of Range ±0.3KVAR ±2% of Range
1000 – 9999 KVAR 1 KVAR ±1% of Range ±3KVAR ±2% of Range

Power Factor (PF)

Range Resolution Accuracy
> 20V & > 30A < 20V or < 30A
0.000 – 1.000 0.001 ± 0.04 ±0.1

AC Current

Range Resolution Accuracy of Readings
3A – 300.0A 0.1A ±1% of Range ±0.3A
300.0 – 3000A 0.1A / 1A ±1% of Range ±3A

AC Voltage (Overload Protection AC 800V)

Range Resolution Accuracy of Readings
4.0 V – 600.0 V 0.1 V ±0.5% ± 5dgts

Harmonics of AC Voltage in Percentage

Range Resolution Accuracy
1 – 20th 0.1% ±2%
21 – 50th 0.1% 4% of reading ±2.0%

Harmonics of AC Voltage in Magnitude

Range Resolution Accuracy
1 – 20th 0.1% ±2% ± 0.5V
21 – 50th 0.1% 4% of reading ±0.5V

Harmonics of AC Current in Percentage

Range Resolution Accuracy
1 – 20th 0.1% ±2%
21 – 50th 0.1% ±6%

Harmonics of AC Current in Magnitude (True RMS < 300A)

Range (0 – 300A) Resolution Accuracy
1 – 20th 0.1% ±2% of range ±0.4A
21 – 50th 0.1% ±4% of range ±0.4A

Harmonics of AC Current in Magnitude (3000A >True RMS> 300A)

Range Resolution Accuracy
1 – 20th 0.1% ±2% of range ±40A
21 – 50th 0.1% ±4% of range ±40A

Phase Angle (Φ, V > 20V, A > 30A)

Range Resolution Accuracy
-180° to 180° 0.1° ± 2°
0° to 360° 0.1° ± 2°

Total Harmonic Distortion

Range Resolution Accuracy
0.0 – 20% 0.1% ± 2%
20 – 100% 0.1% ± 7%
100 – 999.9 % 0.1% ± 11%

Peak Value of AC Voltage (RMS value > 20V) or AC Current (RMS value > 30A)

Range Sampling Time Accuracy of Reading
50 Hz 39 μs ± 5% ± 50 digits
60 Hz 33 μs ± 5% ± 50 digits

Frequency of AC Voltage (RMS value > 20V) or AC Current (RMS value > 30A)

Range Resolution Accuracy
46 – 65 0.1 ± 0.3Hz

 