| SKU |
HI98195-web |
| Product Name |
Multiparameter pH/ORP/EC/Pressure/Temperature Waterproof Meter – HI98195 |
| Quote Required |
Yes |
| pH Range |
0.00 to 14.00 pH |
| pH Resolution |
0.01 pH |
| pH Accuracy |
±0.02 pH |
| pH Calibration |
automatic one, two, or three points with automatic recognition of five standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer |
| ORP Range |
±2000.0 mV |
| ORP Resolution |
0.1 mV |
| ORP Accuracy |
±1.0 mV |
| ORP Calibration |
automatic at one custom point (relative mV) |
| mV Range |
±600.0 mV |
| mV Resolution |
0.1 mV |
| mV Accuracy |
±0.5 mV |
| EC Range |
0 to 9999 µS/cm; 0.000 to 200.0 mS/cm (absolute EC up to 400.0 mS/cm) |
| EC Resolution |
auto-ranging |
| EC Accuracy |
±1% of reading or ±1 µS/cm whichever is greater |
| EC Calibration |
automatic single point, with six standard solutions (84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point |
| TDS Range |
0 to 9999 ppm (mg/L); 0.000 to 400.0 ppt (g/L) (maximum value depends on TDS factor) |
| TDS Resolution |
manual |
| TDS Accuracy |
±1% of reading or ±1 ppm (mg/L), whichever is greater |
| TDS Calibration |
based on conductivity calibration |
| Salinity (PSU) Range |
0.00 to 70.00 PSU |
| Salinity (PSU) Resolution |
0.01 PSU |
| Salinity (PSU) Accuracy |
±2% of reading or ±0.01 PSU whichever is greater |
| Salinity (PSU) Calibration |
based on conductivity calibration |
| Salinity (Seawater Sigma) Range |
0.0 to 50.0 σt, σ₀, σ₁₅ |
| Salinity (Seawater Sigma) Resolution |
0.1 σt, σ₀, σ₁₅ |
| Salinity (Seawater Sigma) Accuracy |
±1 σt, σ₀, σ₁₅ |
| Salinity (Seawater Sigma) Calibration |
based on conductivity calibration |
| Resistivity Range |
0 to 999999 Ω•cm; 0 to 1000.0 kΩ•cm; 0 to 1.0000 MΩ•cm |
| Resistivity Resolution |
dependent on resistivity reading |
| Resistivity Calibration |
based on conductivity calibration |
| Temperature Range |
-5.00 to 55.00 °C, 23.00 to 131.00 °F, 268.15 to 328.15 K |
| Temperature Resolution |
0.01 K, 0.01 °C, 0.01 °F |
| Temperature Accuracy |
±0.15 °C; ±0.27 °F; ±0.15 K |
| Temperature Calibration |
automatic at one custom point |
| Temperature Compensation |
automatic from -5 to 55 °C (23 to 131 °F) |
| Logging Memory |
45,000 records (continuous logging or log-on-demand of all parameters) |
| Logging Interval |
one second to three hours |
| Connectivity |
USB to PC with HI9298194 Hanna software installed |
| Battery Type/Life |
1.5V AA batteries (4) / approximately 360 hours of continuous use without backlight (50 hours with backlight) |
| Environment |
0 to 50°C (32 to 122°F); RH 100% (IP67) |
| Dimensions |
185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”) |
| Weight |
400 g (14.2 oz.) |
| Ordering Information |
HI98195 is supplied with HI7698195 probe, HI7698194-1 pH/ORP sensor, HI7698194-3 EC sensor, HI 76981952 Probe maintenance kit (O-rings for sensors (5), syringe with grease to lubricate the O-rings), HI9298194 Hanna PC software, HI920015 micro USB cable, 1.5V AA batteries (4), instruction manual, quick start guide, quality certificate, and rugged carrying case with custom thermoformed insert. |
| Atmospheric Pressure Range |
450 to 850 mm Hg; 17.72 to 33.46 in Hg; 600.0 to 1133.2 mbar; 8.702 to 16.436 psi; 0.5921 to 1.1184 atm; 60.00 to 113.32 kPa |
| Atmospheric Pressure Resolution |
0.1 mm Hg; 0.01 in Hg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa |
| Atmospheric Pressure Accuracy |
±3 mm Hg within ±15% from calibration point |
| Atmospheric Pressure Calibration |
automatic at one custom point |