Hanna Instruments – HI2040 edge® Multiparameter DO Meter

The HI2040 edge® Multiparameter DO Meter is a sleek, high-performance laboratory meter designed for accurate dissolved oxygen, pH, EC, TDS, salinity, and temperature measurements. It features digital smart electrodes with built-in calibration memory for seamless plug-and-play operation. Supplied with the HI764080 Clark-type polarographic DO probe, it ensures precise oxygen analysis with automatic temperature compensation. With dual USB connectivity, wall or bench mounting options, and up to 1000 data logs, edge® combines innovation, portability, and precision.

  • Digital Smart Electrodes: Calibration data stored in electrode microchip
  • Clark-Type Polarographic DO Probe: Accurate dissolved oxygen measurement with replaceable PTFE membrane
  • Advanced Data Logging: Up to 1000 records with GLP data
  • Compact & Adaptable Design: Wall mount or cradle use, 8-hour rechargeable battery

Description

The HI2040 edge® DO Meter integrates advanced digital electrode technology featuring a built-in microprocessor that retains calibration data and automatically transfers it to the meter upon connection. The 3.5 mm probe input ensures quick, alignment-free connection without risk of bent pins. Included with the meter is the HI764080 Clark-type polarographic dissolved oxygen probe, designed with a slim 12″ PEI body and integrated temperature sensor for automatic temperature correction. The probe uses a PTFE screw-on membrane cap for easy replacement and operates with HI7041S electrolyte fill solution. By applying voltage between a platinum cathode and silver anode, the probe generates a current proportional to dissolved oxygen concentration for highly reliable measurements.

Designed for flexibility, edge® can be wall-mounted for zero bench footprint or placed in a tabletop cradle with adjustable electrode holder for optimal viewing while charging. The meter offers manual, stability, and interval logging up to 1000 records, storing readings, GLP data, date, and time. With support for dissolved oxygen, pH, EC, TDS, and salinity measurements, along with automatic temperature, salinity, and altitude compensation, the HI2040 delivers laboratory-grade accuracy in a compact and portable format.

Features

  • SKU: HI2040-02
  • Product Name: edge® Multiparameter DO Meter – HI2040
  • Quote Required: Yes
  • Digital Electrodes: Built-in microprocessor storing calibration data for automatic recognition
  • Probe Input: 3.5 mm connector for easy, alignment-free attachment
  • Included DO Probe: HI764080 Clark-type polarographic probe with integrated temperature sensor
  • Membrane Type: PTFE screw-on replacement caps
  • Dissolved Oxygen Range: 0.00 to 45.00 ppm (mg/L); 0.0 to 300.0% saturation
  • Dissolved Oxygen Accuracy: ±1 digit; ±1.5% of reading
  • Dissolved Oxygen Calibration: 1 or 2 points (0% with HI7040; 100% in air)
  • Dissolved Oxygen Compensation: Automatic temperature, salinity (0–40 g/L), altitude (-500 to 4000 m)
  • pH Range: -2.000 to 16.000 pH
  • pH Accuracy: ±0.002 pH
  • pH Calibration: Automatic up to 3 points (5 points in Standard Mode)
  • pH Diagnostics: CAL Check™ including probe condition and response time
  • EC Range: 0.00 μS/cm to 500.0 mS/cm
  • EC Accuracy: ±1% of reading
  • TDS Range: Up to 400.0 g/L
  • Salinity Range: 0.0 to 400.0% NaCl; 2.00 to 42.00 PSU; 0.0 to 80.0 g/L
  • Temperature Range: -20.0 to 120.0 °C (-4.0 to 248.0 °F)
  • Temperature Accuracy: ±0.5 °C
  • Logging Types: Log-on-demand (max 200); log-on-stability (max 200); interval logging (max 600 samples, 100 lots)
  • Logging Memory: Up to 1000 records with GLP data
  • Connectivity: 1 micro USB (charging & PC); 1 USB (flash storage)
  • Battery Life: Built-in rechargeable battery up to 8 hours continuous use
  • Dimensions: 202 x 140 x 12.7 mm
  • Weight: 250 g (8.82 oz)
  • Included Accessories: Benchtop docking station, wall mount cradle, HI764080 DO probe, USB cable, 5 VDC adapter, instruction manual

Specifications

SKU HI2040-02
Product Name edge® Multiparameter DO Meter – HI2040
Quote Required Yes
pH Range -2.000 to 16.000 pH, -2.00 to 16.00 pH, ±1000 mV
pH Resolution 0.001 pH, 0.01 pH, 0.1 mV
pH Accuracy ±0.002 pH, ±0.01 pH, ±0.2 mV
pH Calibration Automatic, up to 3 points (5 points*) calibration, 5 standard (7 standard*) buffers available (1.68*, 4.01 or 3.00, 6.86, 7.01, 9.18, 10.01, 12.45*) and 2 custom buffers*
pH Temperature Compensation ATC
pH CAL Check (electrode diagnostics) glass and reference junction diagnostics ( HI 11311 & HI 12301 only), out of calibration range , probe condition, response time
EC Range 0.00 to 29.99 μS/cm, 3.00 to 29.99 μS/cm, 30.0 to 299.9 μS/cm, 300 to 2999 μS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm, up to 500.0 mS/cm, absolute conductivity ‡
EC Resolution 0.01 μS/cm, 00.1 μS/cm, 0.01 μS/cm, 0.01 mS/cm, 00.1 mS/cm
EC Accuracy ±1% of reading (±0.05 μS/cm or 1 digit, whichever is greater)
EC Calibration 1 Cell Factor calibration; 6 standards available
TDS Range 0.00 to 14.99 ppm (mg/L), 15.0 to 149.9 ppm (mg/L), 150. to 1499. ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L, up to 400.0 g/L absolute TDS ‡ (with 0.80 factor)
TDS Resolution 0.01 ppm, 0.1 ppm, 1 ppm, 0.01 g/L, 0.1 g/L
TDS Accuracy ±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
TDS Calibration through EC calibration
EC/TDS Temperature Compensation Automatic -5 to 100 ºC (23 to 212 ºF); NoTC – can be selected to measure absolute conductivity
Temperature Correction Coefficient 0.00 to 6.00% / ºC (for EC and TDS only). Default value is 1.90% / ºC
EC to TDS Conversion Factor 0.40 to 0.80 (default value is 0.50)
Salinity (% NaCl) Range 0.0 to 400.0% NaCl *
Salinity (% NaCl) Resolution 0.1% NaCl
Salinity (% NaCl) Accuracy ±1% of reading (@25ºC/77ºF)
Salinity (% NaCl) Calibration 1 point with HI 7037L 100% NaCl sea water standard
Salinity (PSU) Range 2.00 to 42.00 PSU *
Salinity (PSU) Resolution 0.01 PSU
Salinity (PSU) Accuracy ±1% of reading (@25ºC/77ºF)
Salinity (PSU) Calibration through EC calibration
Salinity Range 0.0 to 80.0 g/L *
Salinity Resolution (ppt) 0.01 g/L
Salinity Accuracy (ppt) ±1% of reading (@25ºC/77ºF)
Salinity Calibration (ppt) through EC calibration
Dissolved Oxygen Range 0.00 to 45.00 ppm (mg/L), 0.0 up to 300.0% saturation
Dissolved Oxygen Resolution 0.01 ppm, 0.1% saturation
Dissolved Oxygen Accuracy ±1 digit, ±1.5% of reading
Dissolved Oxygen Calibration one or two points at 0% (HI 7040 solution) and 100% (in air)
Dissolved Oxygen Temperature Compensation ATC
Salinity Compensation 0 to 40 g/L (with 1 g/L resolution)
Altitude Compensation -500 to 4000 m (with 100 m resolution)
Temperature Range -20.0 to 120.0 ºC, -4.0 to 248.0 °F
Temperature Resolution 0.1 °C, 0.1 °F
Temperature Accuracy ±0.5 °C, ±1.0 °F
DO Probe HI 764080 (included)
Logging Type Manual log-on-demand (Max. 200 logs); Manual log-on-stability (Max. 200 logs); Interval logging *** (Max. 600 samples; 100 lots)
Logging Memory up to 1000 records
Connectivity 1 micro USB port for charging and PC connectivity, 1 USB port for storage
Battery Type/Life Built-in rechargeable battery with up to 8 hours of continuous use
Power Supply 5 VDC adapter (included)
Environment 0 to 50°C (32 to 122°F), RH max 95% non-condensing
Dimensions 202 x 140 x 12.7mm (8” x 5.5” x 0.5”)
Weight 250 g (8.82 oz)
Ordering Information HI2040 includes
Notes * standard Mode Only
** temperature limits will be reduced to actual probe/sensor limits.