Description
| SKU | HI981037 |
|---|---|
| Product Name | Skin and Scalp pH Tester – HI981037 |
| Quote Required | Yes |
| pH Range | 0.00 to 12.0 pH |
| pH Resolution | 0.01pH |
| pH Accuracy | ±0.05 pH @ 25 °C / 77 °F |
| pH Calibration | automatic one or two-point calibration |
| pH Electrode | pH electrode with flat tip and open junction for surface measurements |
| Automatic Shut-Off | user selectable |
| Battery Type/Life | CR2032 lithium ion battery (1) / approximately 1000 hours of continuous use |
| Environment | 0 to 50°C (32 to 122°F); RH 95% max |
| Dimensions | 50 x 151 x 21 mm (2 x 5.9 x 0.9”) |
| Weight | 44 g (1.6 oz.) |
| Ordering Information | HI981037 is supplied with pH 4.01 buffer solution sachet (2), pH 7.01 buffer solution sachet (2), cleaning and disinfection solution sachet for skin residuals, electrode cleaning solution sachet for skin grease and sebum, electrode storage solution (13 mL), CR2032 3V Li-ion battery, quality certificate, and instruction manual. |
Cleaning and Disinfection Solution
Your Skin pH Tester comes with all the required solution like cleaning and disinfection solution for skin residual.
pH 4 and pH 7
When performing a pH calibration it is important that the correct buffers are used. We include some start buffers to get you going. The single-use sachets are individually sealed so you use fresh buffer each time you calibrate.
Storage Solution
Proper storage of your electrode will ensure reliable measurements and a long electrode life. Our storage solution is specially formulated to keep the pH glass hydrated while minimizing organic growth.
Features at-a-glance
Large LCD
- Displays the measurement reading, stability indicator, low battery indicator, and calibration tags.
Automatic Calibration
- This pH tester is calibrated automatically to one or two points. Buffers are recognized automatically and after calibration, buffer values used are shown on the display.
- An hourglass indicator is displayed on the LCD until a stable reading is obtained.
Automatic Shut-off
- The meter can be set to automatically turn off after 8 minutes or 60 minutes to conserve battery life in the event that the meter is left on. The auto-off feature can also be disabled.
Probe Diagnostic
- During calibration the meter will display error (Err) message as an indicator that the probe needs to be cleaned.
Low Battery Indicator
- The skin & scalp pH tester has an exceptional battery life of approximately 1000 hours. When the battery power is running low, a low battery indicator is displayed.