Description
Electrical Specifications:(23+/- 5℃, 10 minutes after turning on the power)
DCmA SOURCE (Max. load 1K, Max. voltage supply 24V)
| Range | Resolution | Accuracy |
| 4 – 20mA, 0 – 20mA, 0 – 24 mA | 1µA | ± 0.05% ±8µA |
When output open, the LCD displays “OL”.
DCmV, DCV SOURCE (Max. load 1mA, Short circuit protection<100mA)
| Range | Resolution | Accuracy |
| 0 – 100mV | 10µV | ± 0.05% ± 30µV |
| 0.1 – 1.0000V | 100µV | ± 0.05% ± 300µV |
| 1 – 12.000V | 1mV | ± 0.05% ± 3mV |
When output short circuit, the LCD displays “OL”.
ACmA MEASURE (Ture RMS)
| Range | Resolution | Accuracy | |
| 50/60Hz | 40-1KHz | ||
| 24mA | 1µA | ±1%±5dgts | ±2%±5dgts |
DCmA MEASURE
| Range | Resolution | Accuracy |
| 24mA | 1µA | ±0.05%±2dgts |
DCV MEASURE (Overload protection AC 600V, Input impedance 10MΩ)
| Range | Resolution | Accuracy |
| 2.4V | 0.1 mV |
±1%±3dgts |
| 24V | 1 mV | |
| 240V | 10 mV | |
| 600V | 100mV | ±1.5%±3dgts |
ACV MEASURE (Ture RMS)
(Overload protection AC 600V, Input impedance 10MΩ)
| Range | Resolution | Accuracy | |
| 50/60Hz | 40-1KHz | ||
| 24V | 1 mV | ±1%±5dgts |
±2%±5dgts |
| 240V | 10 mV | ±1%±5dgts | |
| 600V | 100mV | ±1.5%±5dgts | |
Frequency SOURCE (TTL, Square wave)
| Range (Hz) | Resolution | Accuracy |
| 1.0~1000.0Hz | 0.1Hz | 0.1Hz |
| 1000~10000Hz | 1Hz | 1Hz |
| 10000~20000Hz | 1Hz | 10Hz |
Frequency MEASURE (Sensitivity RMS 1V)
| Range (Hz) | Resolution | Accuracy |
| 1.000~39.999Hz | 0.001Hz | ±0.5%±0.003Hz |
| 40.00~399.99Hz | 0.01Hz | ±0.5%±0.03Hz |
| 0.4000K~3.999KHz | 0.1Hz | ±0.5%±0.3Hz |
| 4.000K~39.999KHz | 1Hz | ±0.5%±3Hz |
| 40.00K~399.99KHz | 10Hz | ±0.5%±30Hz |
| 0.4000M~3.999MHz | 1KHz | ±0.5%±3KHz |
| 4.000M~39.999MHz | 1KHz | ±0.5%±3KHz |
Duty Cycle SOURCE/MEASURE (1~20KHz)
| Range | Resolution | Accuracy |
| 1% to 99% | 1% | ±1%±30μS |
Temperature Thermocouples
(SOURCE and MEASURE, 0.1C & 0.1F Resolution, Internal Cold Junction Compensation, thermocouples accuracy not included, 3 minutes after plugging in thermocouples.)
| °C | °F | |||
| Range | Accuracy | Range | Accuracy | |
| K | -200 to -150 | 2.0 | -382 to -238 | 3.6 |
| -150 to 0 | 1.2 | -238 to 32 | 2.1 | |
| 0 to 1000 | 0.8 | 32 to 1832 | 1.4 | |
| 1000 to 1370 | 1.2 | 1832 to 2498 | 2.1 | |
| J | -200 to -150 | 2.0 | -382 to -238 | 3.6 |
| -150 to 0 | 1.0 | -238 to 32 | 1.8 | |
| 0 to 1050 | 0.7 | 32 to 1922 | 1.2 | |
| E | -200 to -150 | 1.5 | -382 to -238 | 2.7 |
| -150 to 0 | 0.9 | -238 to 32 | 1.6 | |
| 0 to 850 | 0.7 | 32 to 1562 | 1.2 | |
| T | -200 to -150 | 1.5 | -382 to -238 | 2.7 |
| -150 to 0 | 1.2 | -238 to 32 | 2.1 | |
| 0 to 400 | 0.8 | 32 to 752 | 1.4 | |
| R | 0 to 500 | 1.8 | 32 to 932 | 3.2 |
| 500 to 1760 | 1.5 | 932 to 3200 | 2.7 | |
| S | 0 to 500 | 1.8 | 32 to 932 | 3.2 |
| 500 to 1760 | 1.5 | 932 to 3200 | 2.7 | |
| N | -200 to 0 | 1.5 | -328 to 32 | 2.7 |
| 0 to 1300 | 0.9 | 32 to 2372 | 1.6 | |
| L | -200 to 0 | 0.9 | -328 to 32 | 1.6 |
| 0 to 900 | 0.7 | 32 to 1652 | 1.2 | |
| U | -200 to 0 | 1.1 | -328 to 32 | 1.9 |
| 0 to 600 | 0.7 | 32 to 1112 | 1.2 | |
| B | 600 to 800 | 2.2 | 1112 to 1472 | 3.9 |
| 800 to 1000 | 1.8 | 1472 to 1832 | 3.2 | |
| 1000 to 1820 | 1.4 | 1832 to 3308 | 2.5 | |
| C | 0 to 1800 | 1.0 | 32 to 3272 | 1.8 |
| 1800 to 2310 | 1.5 | 3272 to 4190 | 2.7 | |
Resistance (Ω) MEASURE (Open voltage 0.4V, Overload protection AC 600V)
| Range | Resolution | Accuracy |
| 400Ω | 0.01Ω |
±1%±2dgts |
| 4KΩ | 0.1Ω | |
| 40KΩ | 1Ω | |
| 400KΩ | 10Ω | |
| 4MΩ | 100Ω |
±1.5%±2dgts |
| 40MΩ | 1KΩ |
Continuity MEASURE:
Open voltage 2V, Overload protection AC 600V, <10Ω Beep.
Diode MEASURE (Open voltage 2V, Overload protection AC 600V)
| Range | Resolution | Accuracy |
| 0~1.9999V | 0.0001V | ±2.5%±5dgts |
General Specifications:
| Dimension | 214.0 (L) x 98.7(W) x 56.0(H) mm
8.4″ (L) x 3.9″ (W) x 2.2″ (H) |
| Weight | 650g / 22.9oz (batteries included) |
| Operation Environment | 0℃ ~ 50℃
< 85% RH |
| Storage Environment | -20℃ ~ 60℃ , <75% RH |
| Accessories | Carrying case x 1 User manual x 1
1.5V SUM-3 Battery x 5 K-type thermocouples (dual plugs) x 1 Alligator clips x 2 (black and red) Test leads x 2 (black and red) |