Phone

+603-5621 5786

Business Hour

08:30 - 17:30

Met One- SASS and SUPER SASS Speciation Sampler

The Met One SASS and SUPER SASS are advanced PM2.5 speciation samplers designed to measure particulate mass, trace metals, organic and elemental carbon, sulfates, nitrates, and other ions. They provide reliable, multi-day ambient air sampling with precise flow control and integrated data logging.

  • Comprehensive PM2.5 speciation: Measures mass, metals, carbon, and ions simultaneously
  • Multi-canister operation: SASS supports 5 canisters, SUPER SASS up to 8 for extended sampling
  • Automated volumetric flow control: Each channel operates independently with accurate flow rates
  • Portable and field-ready: Includes tripod stand, solar shields, and NEMA environmental enclosure for robust outdoor use

Description

The Met One SASS and SUPER SASS Speciation Samplers are advanced, portable systems designed for high-precision ambient particulate monitoring, focusing on PM2.5 mass, trace metals, organic and elemental carbon, sulfates, nitrates, and other ions. The SASS system accommodates up to five canisters, while the SUPER SASS supports up to eight, each with active flow-controlled channels at 6.7–6.9 L/min per canister. Both systems utilize sharp cut cyclones (D50 2.5 µm) and multicell denuders with magnesium oxide, sodium carbonate, or citric coatings, along with 47 mm FRM filters, ensuring high-quality, contaminant-free sample collection. Temperature, volumetric flow, start/stop times, and sampler status are logged at five-minute intervals, and solar-protected, wind-aspirated shields maintain filter temperatures within 5°C above ambient, preserving sample integrity.

The SUPER SASS expands on the SASS functionality with sequential, multi-day sampling capability, allowing for continuous operation of up to eight days. It features four separate volumetric flow controllers for autonomous channel operation, enabling precise, independent management of multiple sample streams. Both systems are mounted on tripods with environmental control boxes, utilizing two-head diaphragm vacuum pumps powered by 110 V AC and a 12 V DC fan. These samplers are ideal for field audits, environmental research, and regulatory monitoring, providing reliable, high-resolution speciation data for PM2.5 constituents in both short-term and extended sampling campaigns.

 

Features

SASS FEATURES

  • Integrated portable sampling equipment for ambient particulates.
  • 6.7 litres per minute PM2.5 inlet sampling rate.
  • Cassettes are kept at less than 5°C above ambient temperature by the solar protection.
  • Data integrity is provided by canister, which is contaminant-proof.
  • Various filter medium and a new multi-cell denuder.

SUPER SASS FEATURES

This device has all of the functionality of the SASS in addition to having sequential programming for multi-day operation. Depending on the number of sample channels utilised each day, SUPER SASS will sample for up to eight days. Four separate volumetric flow controllers, one for each sampling group, are included with SUPER SASS as standard.

  • 8 channel operation maximum.
  • Sequential sampling for one to eight days.
  • Controllers for Automatic Volumetric Flow.
  • Each Channel is capable of autonomous operation.
  • Screens for Advanced Field Audits.

Specifications

Analytes: PM2.5 mass and trace metals

PM2.5 organic and elemental carbon

PM2.5 sulfate, nitrate and other ions

PM2.5 elements

Number of Canisters: SASS = Five (5), SuperSASS = Eight (8) maximum
Sample Flow Rate: 6.7 Liters/minute ± 2% per canister (active flow controlled channels

6.9 Liters/minute per canister (critical orifice channels

Inlet: Sharp cut cyclone (SCC, (D50 2.5µm AED Detachable from Canister
Denuders: Multicell, magnesium oxide, sodium carbonate, citric

 

Filter size: 47 mm OD typical FRM filter
Filters per One or two in series
canister Plenum: None
Transport Lines None
Logged Parameters Ambient temperature, °C (-30 to 50

Filter temperature, °C (-30 to 50

Sample start date and time,

Sample stop date and time

Status of Sampler, Flags

Volumetric flow rates (L/m, and Volume (m3/hr

Logging Interval 5 minute averaging
Support Stand Tripod with mounting feet and mast
Solar radiation Shield, 20” Diameter. x 12”H, Wind Aspirated
Control Box NEMA Environmental Enclosure 14”L x 14”W x 19.5”H including mounting base.
Pump Two head diaphragm vacuum, 110/60AC
Power 110AC/60 Hz input to pump system and fan 12 VDC
Ambient Temp (-30 to 50°C Bead thermocouple in naturally aspirated radiation shield
Filter Temp (-30 to 50°C Bead thermocouple at inlet of each canister