Description
The handheld process calibrator allows technicians to source and measure electrical signals efficiently. It can generate 4–20 mA and 0–24 mA currents with 1 µA precision and 0–24 V DC voltages with 1 mV resolution. The device produces square-wave frequencies from 1 Hz to 20,000 Hz with a fully programmable duty cycle of 1–99%. These features make it suitable for testing industrial meters, panel instruments, and LED brightness.
Technicians can measure DC mA, DC V, and temperature at the same time. It supports 11 types of thermocouples, offering 0.1 °C / 0.1 °F resolution and internal cold junction compensation. The calibrator alerts users to open outputs, overloads, and short circuits, and shows battery percentage on the LCD. Its compact size (214 × 98.7 × 56 mm, 650 g) and included accessories (thermocouples, alligator clips, leads, batteries) make it ready for field and laboratory work.
Specifications
General Specifications:
| Dimension |
214.0 (L) x 98.7(W) x 56.0(H) mm
8.4″ (L) x 3.9″ (W) x 2.2″ (H) |
| Weight |
650g / 22.9oz (batteries included) |
| Operation Environment |
0°C ~ 50°C
< 85% RH |
| Storage Environment |
-20°C ~ 60°C
< 75% RH |
| Accessories |
Carrying case x 1 User manual x 1
1.5V SUM-3 Battery x 5
K-type thermocouples (1M, single plug) x 1 Alligator clips x 2 (black and red)
Test leads x 2 (black and red) |
Electrical Specifications:
(23+/- 5°C, 10 minutes after turning on the power)
DC mA SOURCE (Max. load 1KΩ, Max. voltage supply 24V)
| Range |
Resolution |
Accuracy |
| 0~4mA |
1μA |
± 0.025% ±10μA |
| 4~20mA |
1μA |
± 0.025% ±5μA |
| 20~24mA |
1μA |
± 0.025% ±5μA |
When output Open, the LCD displays “OL”.
DC V SOURCE (Max. load 24mA, Short circuit protection < 25mA)
| Range |
Resolution |
Accuracy |
| 0~4V |
1mV |
± 0.05% ± 10mV |
| 4~20V |
1mV |
± 0.05% ±5mV |
| 20~24V |
1mV |
± 0.05% ±5mV |
Frequency SOURCE (TTL, Square wave, Duty cycle = 1~99%)
| Range(Hz) |
Resolution |
Accuracy |
| 1.0~1000.0Hz |
0.1Hz |
0.1Hz |
| 1000~10000Hz |
1Hz |
1Hz |
| 10000~20000Hz |
1Hz |
10Hz |
DC mA MEASURE
| Range |
Resolution |
Accuracy |
| 0.005~4mA |
1μA |
±0.025%±5dgts |
| 4~20mA |
1μA |
±0.025%±5dgts |
| 20~24mA |
1μA |
±0.025%±5dgts |
If reading of mA (measure) is less than 5 digits, it is displayed as 0.
DC V MEASURE
| Range |
Resolution |
Accuracy |
| 0~24V |
0.001V |
±0.025%±5dgts |
If reading of V (measure) is less than 5 digits, it is displayed as 0.
Temperature of Thermocouples (MEASURE)
(Resolution 0.1C&0.1F, Internal cold junction compensation, Thermocouples accuracy not included, 3 minutes after plugging in thermocouples.)
|
°C |
°F |
| Range |
Accuracy |
Range |
Accuracy |
| K |
-200 to -150 |
2.0 |
-382 to -238 |
3.6 |
| -150 to 0 |
1.2 |
-238 to 32 |
2.1 |
| 0 to 1000 |
0.8 |
32 to 1832 |
1.4 |
| 1000 to 1370 |
1.2 |
1832 to 2498 |
2.1 |
| J |
-200 to -150 |
2.0 |
-382 to -238 |
3.6 |
| -150 to 0 |
1.0 |
-238 to 32 |
1.8 |
| 0 to 1050 |
0.7 |
32 to 1922 |
1.2 |
| E |
-200 to -150 |
1.5 |
-382 to -238 |
2.7 |
| -150 to 0 |
0.9 |
-238 to 32 |
1.6 |
| 0 to 850 |
0.7 |
32 to 1562 |
1.2 |
| T |
-200 to -150 |
1.5 |
-382 to -238 |
2.7 |
| -150 to 0 |
1.2 |
-238 to 32 |
2.1 |
| 0 to 400 |
0.8 |
32 to 752 |
1.4 |
| R |
0 to 500 |
1.8 |
32 to 932 |
3.2 |
| 500 to 1760 |
1.5 |
932 to 3200 |
2.7 |
| S |
0 to 500 |
1.8 |
32 to 932 |
3.2 |
| 500 to 1760 |
1.5 |
932 to 3200 |
2.7 |
| N |
-200 to 0 |
1.5 |
-328 to 32 |
2.7 |
| 0 to 1300 |
0.9 |
32 to 2372 |
1.6 |
| L |
-200 to 0 |
0.9 |
-328 to 32 |
1.6 |
| 0 to 900 |
0.7 |
32 to 1652 |
1.2 |
| U |
-200 to 0 |
1.1 |
-328 to 32 |
1.9 |
| 0 to 600 |
0.7 |
32 to 1112 |
1.2 |
| B |
600 to 800 |
2.2 |
1112 to 1472 |
3.9 |
| 800 to 1000 |
1.8 |
1472 to 1832 |
3.2 |
| 1000 to 1820 |
1.4 |
1832 to 3308 |
2.5 |
| C |
0 to 1800 |
1.0 |
32 to 3272 |
1.8 |
| 1800 to 2310 |
1.5 |
3272 to 4190 |
2.7 |