Description
The multifunction calibrator provides precise sourcing and measurement of mA, V, Hz, and temperature. It can source 4–20 mA, 0–20 mA, and 0–24 mA signals with 1 μA resolution. The device measures DC voltage and current with high accuracy. Users can simulate electronic loads and map signals into engineering units for easy calibration.
It includes a handheld function generator for sine, square, triangular, and arbitrary waves from 0.3 Hz to 20 kHz. Pulse generation ranges from 3 μS to 999.99 mS. The calibrator can measure temperature using 11 thermocouple types with programmable cold junction compensation. Additional features include DTMF testing, backlit LCD display, rechargeable battery, data logging, and USB programmability, making it ideal for field or laboratory applications.
Specifications
General Specifications:
| AC Power Adaptor: |
AC 110V or AC 220V, 50/60Hz input.
DC 15V / 0.5A output |
| Dimension: |
214.0 (L) x 98.7 (W) x 56.0 (H) mm
8.4″ (L) x 3.9″ (W) x 2.2″ (H) |
| Weight: |
650g / 22.9oz (Batteries included) |
| Operation Environment: |
0℃ ~ 50℃, <85% RH |
| Storage Environment: |
-20℃ ~ 60℃, <75% RH |
| Accessories: |
Carrying case x 1, User manual x 1, AC power adaptor x 1, USB cable x 1, Software CD x 1, Software manual x 1, K-type thermocouple (dual plugs) x 1, Alligator clips x 2 (black and red),
Test leads x 2 (black and red),
Rechargeable lithium battery (11.1V/ 1600mAh) x 1 |
Electrical Specifications: (23+/- 5℃, 10 minutes after turning on the power)
mA (source) (Vopen > 24V)
| Range |
Resolution |
Accuracy of Reading |
| 0.005mA to 4mA |
1uA |
+/-0.03% +/-5dgts |
| 4mA to 20mA |
+/-0.03% +/-3dgts |
| 20mA to 24mA |
+/-0.03% +/-5dgts |
V (source) (maximum load 1mA, short circuit protection < 100mA)
| Range |
Resolution |
Accuracy of Reading |
| 0.005V to 10V |
0.001V |
+/-0.03% +/-5dgts |
| 10V to 15V |
mA (measure)
| Range |
Resolution |
Accuracy of Reading |
| -4mA to -0.005mA |
1μA |
+/-0.03% +/-10dgts |
| 0.005mA to 4mA |
+/-0.03% +/-5dgts |
| 4mA to 20mA |
+/-0.03% +/-3dgts |
| 20mA to 24mA |
+/-0.03% +/-5dgts |
If reading of mA (measure) is less than 5 digits, it is displayed as 0.
V (measure)
| Range |
Resolution |
Accuracy of Reading |
| -3V to -0.005V |
0.001V |
+/-0.03% +/-10dgts |
| 0.005V to 10V |
+/-0.03% +/-5dgts |
| 10V to 24V |
+/-0.03% +/-5dgts |
If reading of V (measure) is less than 5 digits, it is displayed as 0.
Frequency (source, 10 Vpp, 0V offset, square wave, duty cycle = 50%)
| Range (Hz) |
Input Resolution |
Accuracy |
| 0.3 to 99.999 |
0.1Hz |
0.002Hz |
| 10.00 to 999.99 |
0.1Hz |
0.02Hz |
| 1000.0 to 9999.9 |
0.1Hz |
0.2Hz |
| 10000 to 20000 |
1Hz |
2Hz |
Duty Cycle(%, square wave, 10 Vpp, 0.3~20KHz)
| Range |
Resolution |
Rise Time of Vpp |
Fall Time of Vpp |
| 0 to 100% |
1% |
10μS max, 5μS typical |
15μS max, 7.5μS typical |
Pulse (square wave, 10 Vpp, Offset -5V~+5V)
| Range |
Resolution |
Rise Time of Vpp |
Fall Time of Vpp |
| 3.0μS to 9999.9μS |
0.1μS |
10μS max, 5μS typical |
15μS max, 7.5μS typical |
| 10.000mS to 99.999mS |
0.001mS |
| 100.00mS to 999.99mS |
0.01mS |
Voltage Peak to Peak for Sine Wave
(Vpp, 0.3~20KHz, 50% duty cycle, sine wave, 0V offset)
| Range (V) |
Resolution |
Accuracy of Reading |
| 0.1 to 20V |
0.001V |
5% +/- 0.3V |
Voltage Peak to Peak for Non-Sine Wave (Vpp, 0.3~20KHz, 0V offset)
| Range (V) |
Resolution |
Accuracy of Reading |
| 0.1 to 20V |
0.001V |
6% +/- 0.4V |
Voltage Peak to Peak(Vpp, 0.3~20KHz, 50% duty cycle, square wave, 0V offset)
| Range (V) |
Resolution |
Accuracy of Reading |
| 1 to 20V |
0.001V |
6% +/- 0.4V |
Voltage of Offset(Maximum Vpp < 20V)
| Range |
Resolution |
Accuracy of Reading |
| -5V to 5V |
0.001V |
5% +/-0.5V +/-5%xVpp |
Temperature, Thermocouples
(source and measure, 0.1C & 0.1F Resolution, Internal Cold Junction Compensation, thermocouple accuracy not included, 3 minutes after plugging in thermocouples.)
|
°C |
°F |
| Range |
Accuracy |
Range |
Accuracy |
| K |
-200 to -150 |
2.0 |
-382 to -238 |
3.6 |
| -150 to 0 |
1.2 |
-238 to 32 |
2.1 |
| 0 to 1000 |
0.8 |
32 to 1832 |
1.4 |
| 1000 to 1370 |
1.2 |
1832 to 2498 |
2.1 |
| J |
-200 to -150 |
2.0 |
-382 to -238 |
3.6 |
| -150 to 0 |
1.0 |
-238 to 32 |
1.8 |
| 0 to 1050 |
0.7 |
32 to 1922 |
1.2 |
| E |
-200 to -150 |
1.5 |
-382 to -238 |
2.7 |
| -150 to 0 |
0.9 |
-238 to 32 |
1.6 |
| 0 to 850 |
0.7 |
32 to 1562 |
1.2 |
| T |
-200 to -150 |
1.5 |
-382 to -238 |
2.7 |
| -150 to 0 |
1.2 |
-238 to 32 |
2.1 |
| 0 to 400 |
0.8 |
32 to 752 |
1.4 |
| R |
0 to 500 |
1.8 |
32 to 932 |
3.2 |
| 500 to 1760 |
1.5 |
932 to 3200 |
2.7 |
| S |
0 to 500 |
1.8 |
32 to 932 |
3.2 |
|
500 to 1760 |
1.5 |
932 to 3200 |
2.7 |
| N |
-200 to 0 |
1.5 |
-328 to 32 |
2.7 |
| 0 to 1300 |
0.9 |
32 to 2372 |
1.6 |
| L |
-200 to 0 |
0.9 |
-328 to 32 |
1.6 |
| 0 to 900 |
0.7 |
32 to 1652 |
1.2 |
| U |
-200 to 0 |
1.1 |
-328 to 32 |
1.9 |
| 0 to 600 |
0.7 |
32 to 1112 |
1.2 |
| B |
600 to 800 |
2.2 |
1112 to 1472 |
3.9 |
| 800 to 1000 |
1.8 |
1472 to 1832 |
3.2 |
| 1000 to 1820 |
1.4 |
1832 to 3308 |
2.5 |
| C |
0 to 1800 |
1.0 |
32 to 3272 |
1.8 |
| 1800 to 2310 |
1.5 |
3272 to 4190 |
2.7 |
| mV |
-10mV to 70mV |
0.05mV |
-10mV to 70mV |
0.05mV |
DTMF (Hz)
| Range (Hz) |
Resolution |
Accuracy of Reading |
| 0.3 to 99.999 |
0.1Hz |
0.002Hz |
| 10.00 to 999.99 |
0.1Hz |
0.02Hz |
| 1000.0 to 9999.9 |
0.1Hz |
0.2Hz |
| 10000 to 20000 |
1Hz |
2Hz |
DTMF (%)
| Range (%) |
Resolution |
Accuracy of Reading |
| 0% ~ 100% |
1% |
5% |
DTMF (Phase Angle)
| Range (°) |
Resolution |
Accuracy of Reading |
| 0 ~ 360 |
1° |
100μS+1° |
DTMF (Vpp, F1=F2,
| Range |
Resolution |
Accuracy of Reading |
| 5V ~ 20V |
0.001V |
10% +/-0.6V |
DTMF (Offset, F1=F2,
| Range |
Resolution |
Accuracy of Reading |
| -5V ~ 5V |
0.001V |
10% +/-0.6V +/-5%xVpp |