Description
This portable process calibrator provides stable and accurate signal testing for industrial work. It measures and sources 4–20 mA loops, DC voltage, thermocouples, and frequency signals. With ±0.025% mA accuracy and 1 μA resolution, it helps technicians verify transmitters, PLC inputs, and controllers with confidence.
The unit includes safety warning beepers for open and short circuits. It supports K, J, E, and T thermocouples with 1°C / 1°F resolution across a wide temperature range. The calibrator also generates frequency signals up to 62.5 kHz. Its compact body and light weight make it easy to carry to job sites. A 9V battery and external pack option ensure reliable field operation.
Specifications
General Specifications:
| Dimension: |
88 x 168 x 26 mm
(3.46″ x 6.61″ x 1.03″) |
| Battery Type: |
9V Alkaline Battery |
| Display: |
4 and 5 Digits |
| Power Consumption: |
60mA – 180mA |
| Operating Temperature: |
0C to 50C
(32°F to 122°F) |
| Operating Humidity: |
less than 85% relative |
| Storage Temperature: |
-20C to 60C
(-4°F to 140°F) |
| Storage Humidity: |
less than 85% relative |
| Weight: |
330g / 11.63 oz |
| Accessories: |
Carrying case x 1
Users Manual x 1
External Battery Pack: 1.5V AA Batteries x 6 9V Alkaline Battery x 1
Alligator clip x 1
K type thermocouple connector Test leads x 1 |
Electrical Specifications: (23℃±5℃, 10 minutes after power is on)
mA DC Current: (1KΩMax. Load, 24V Loop Supply)
|
Range |
Resolution |
Accuracy |
| 4 – 20mA, 0 – 20mA, 0 -24 mA |
1μA |
± 0.025% ±3μA |
Beeper warning when output is open and specified output > 1mA
mV, V DC Voltage:(1mA Supply Current)
|
Range |
Resolution |
Accuracy |
| 0 – 100.00mV |
10μV |
± 0.05% ± 30μV |
| 0 – 10.000V |
1mV |
± 0.05% ± 3mV |
| 0 – 1.0000V |
100μV |
± 0.05% ± 300μV |
Beeper warning when output is short and specified voltage output > 10mV
K, J, E, T type Thermocouples:(1°C, 1°F Resolution, 1KΩ Load Min.)
| Range |
Accuracy |
Range |
Accuracy |
| K: -200 to -100 °C |
± 2.0°C |
K: -328 to -148°F |
± 3.6°F |
| K: -100 to 0 °C |
± 1.1°C |
K: -148 to 32°F |
± 1.8°F |
| K: 0 to 1370 °C |
± 0.8°C |
K: 32 to 2400°F |
± 1.5°F |
| J: -100 to 0 °C |
± 0.9°C |
J: -148 to 32°F |
± 1.6°F |
| J: 0 to 760 °C |
± 0.7°C |
J: 32 to 1400°F |
± 1.2°F |
| E: -100 to 0 °C |
± 0.9°C |
E: -148 to 32°F |
± 1.6°F |
| E: 0 to 700 °C |
± 0.7°C |
E: 32 to 1292°F |
± 1.2°F |
| T: -200 to 0 °C |
± 1.0°C |
T: -328 to 32°F |
± 1.8°F |
| T: 0 to 400 °C |
± 0.8°C |
T: 32 to 752°F |
± 1.5°F |
Frequency:(1 – 125Hz, TTL 0 – 5V, 1KΩ Load Min.)
| Range |
Resolution |
Accuracy |
|
|
|
Available Frequencies
(126 – 62500Hz, TTL 0 – 5V, Accuracy ± 0.01% ± 0.04Hz, 1KΩ Load Min):
| 126.00 |
127.03 |
128.07 |
129.13 |
130.20 |
131.30 |
132.41 |
133.54 |
134.12 |
135.28 |
| 136.46 |
137.06 |
138.27 |
139.50 |
140.13 |
141.40 |
142.04 |
143.34 |
144.00 |
145.34 |
| 146.02 |
147.40 |
148.10 |
149.52 |
150.24 |
151.69 |
152.43 |
153.18 |
154.70 |
155.47 |
………………………………………….
16666.66 17241.37 17857.14 18518.51 19230.76 20000.00 20833.33 21739.13 22727.27
23809.52 25000.00 26315.78 27777.77 29411.76 31250.00 33333.33 35714.28 38461.53
41666.66 45454.54 50000.00 55555.55 62500.00