E-FRM-DC Reference Method Particulate Sampler
Applications:
Categories: Climatic/Environment Inspection, Particle Counter
Description
Specifications:
| Measurement Principle | Programmable-event filter sampler for determination of particulate matter concentrations |
| Resulting Data Units | Micrograms per actual or standard cubic meter μg/m3. Calculated from lab filter weights and
actual or standard sampled volumes. |
| Design Specifications | Title 40 CFR, Part 50 Appendix L |
| U.S. EPA Designations | RFPS-0315-221, RFPS-0216-231, RFPS-0316-232 |
| Main Inlet Type | EPA-pattern PM10 size-selective louvered inlet standard (MOI part no. BX-802) |
| PM2.5 Separator Type | Options: BX-808 Very Sharp Cut Cyclone or BX-804 WINS Impactor |
| PM10 Sampling Configuration | Pass-through tube (used in place of cyclone) included for PM10 sampling |
| Other Configurations | Compatible with PM2.5 or PM1 SCC cyclones. TSP inlet available. |
| Filter Type | 47mm PTFE disc filters standard. Other filter material types available. |
| Filter Holder | Accommodates standard EPA-pattern white Delrin filter cassettes. |
| Filter Temperature Control | Filter Temp within 5°C of ambient during sampling and standby. Logged in data. |
| Operating Temperature | -30 to +50°C Ambient |
| Operating Pressure | 400 to 800 mm Hg, Ambient |
| Sample Events | 24 hour samples standard. Programmable start date, time and duration |
| Flow Rate | 16.67 L/min (1 m3/hr.). With flow sensor and real-time actual flow control. |
| Flow Accuracy | ±2% (±.33 L/min) |
| Flow Records | Flow rate, temperature, pressure logged in 5 minute intervals |
| Vacuum Pump | Single-head diaphragm 12 VDC pump |
| Data Collection | RS-232 port (9-pin). USB data port. USB flash drive port. Software supplied. |
| Memory Capacity | 24 sample events |
| Compatible Software | FSCommAQ |
| User Interface | 4.3” graphical touch screen |
| Enclosure Type | Active fan ventilation |
| Deployed Dimensions | 2.0 meters (78.75”) high, 65.5 cm (25.75”) wide, 64.1 cm (25.25”) deep |
| Footprint Dimensions | 65.5 cm (25.75”) wide, 64.1 cm (25.25”) deep |
| Weight | Sampler 33 Lbs (except PM10 inlet head); Stand 10.5 Lbs |
| External Power Supply | Input: 100- 240 VAC, 50/60 Hz, 1.4 A; Output: 12 VDC, 8.5 A (102 W) Maximum |
| E-FRM-DC Power | 12 VDC @ 2.75 A (33 W) Maximum |