The GL7-DCB is a 4-channel DC bridge amplifier module for strain gauge sensors, enabling direct measurement of strain, voltage, and resistance with high precision. It supports built-in bridge elements, excitation power, TEDS, and advanced filtering.

  • Strain gauge support – Works with quarter, half, and full-bridge configurations, including remote sensing.
  • High-precision measurement – Sampling from 10 µs to 1 hr, accuracy ±0.2% FS + 10 με (strain) and ±0.2% FS + 10 µV (voltage).
  • Built-in excitation & bridge – Provides DC voltage or constant current for strain gauges; built-in bridge elements (120 Ω / 350 Ω).
  • Advanced filtering & TEDS – Low-pass and anti-aliasing filters, supports IEEE 1451.4 Class2 TEDS for easy sensor setup.

Description

The Graphtec GL7-DCB is a versatile DC bridge amplifier module designed for use with strain gauge-based sensors, enabling precise measurement of strain, voltage, and resistance. Featuring 4 isolated input channels with simultaneous sampling, the module supports quarter, half, and full bridge configurations, including remote sensing for enhanced accuracy. It provides a wide measurement range for strain (up to 20,000 με), voltage (up to 10 V full scale), and resistance (up to 50 kΩ), with excitation power supplied in both voltage and constant current modes. The module’s built-in RAM can store up to 2 million samples per channel, and its high input impedance, robust isolation, and wide frequency response (DC to 20 kHz) ensure reliable and accurate signal acquisition.

The GL7-DCB also supports IEEE 1451.4 TEDS for automatic sensor identification and setup, simplifying configuration and calibration. High-precision measurements are enabled through low-pass and anti-aliasing filters, while features such as automatic zero adjustment and shunt calibration allow easy maintenance of measurement accuracy. Compact and durable, this module is suitable for dynamic testing applications where accurate strain, voltage, or resistance measurements are critical, making it ideal for evaluating phenomena such as force, pressure, and acceleration with strain gauge sensors.

Features

  • Strain gauge support – Works with quarter, half, and full-bridge configurations, including remote sensing.
  • High-precision measurement – Sampling from 10 µs to 1 hr, accuracy ±0.2% FS + 10 με (strain) and ±0.2% FS + 10 µV (voltage).
  • Built-in excitation & bridge – Provides DC voltage or constant current for strain gauges; built-in bridge elements (120 Ω / 350 Ω).
  • Advanced filtering & TEDS – Low-pass and anti-aliasing filters, supports IEEE 1451.4 Class2 TEDS for easy sensor setup.

Specifications

Item Description
Type of module DC bridge amplifier module – for strain gauge or sensor based on a strain gauge –
Model number GL7-DCB
Number of input channels 4 channels
Input method All channels isolated balanced input, Simultaneous sampling,
D-SUB type connector (9 pins, receptacle)
Sampling speed (interval) 100 k Samples/s to 1 Sample/h (10 μs to 1 hr.)
Built in RAM 2 million samples for each channel
Input type Strain, Voltage, Resistance value (including potentiometer)
Measurement range Strain 500, 1000, 2000, 5000, 10000, 20000 με (με: 10-6 strain)
0.2, 0.25, 0.4, 0.5, 1, 2, 2.5, 4, 5, 10 mV/V
* Available ranges vary by the excitation voltage for the bridge.
Voltage 1, 2, 5, 10, 20, 50, 100, 200, 500 mV, 1, 2, 5 V Full Scale
Resistance 1, 2, 5, 10, 20, 50, 100, 200, 500 Ω, 1, 2, 5, 10, 20, 50 kΩ Full Scale
Measurement accuracy (*1) Strain ±(0.2% of Full Scale + 10 με)
Voltage ±(0.2% of Full Scale + 10 μV)
Resistance ±0.5% of Full Scale
A/D converter Successive Approximation type, 16 bits (effective resolution: 1/40000 of the measuring full range)
Input impedance 10 MΩ ±5%
Gauge ratio 2.0 constant
Supported sensor Strain (*2) Strain gauge
Quarter bridge (single gauge) in 2-, 3- or 4-wire (supports remote sensing in 3- or 4-wire)
Half bridge (dual gauge) in 3-, 4-, 5-wire (supports remote sensing in 4- or 5-wire)
Full bridge (quad gauge) in 4- or 6-wire (supports remote sensing in 6-wire)
Transducer/sensor based on a strain gauge
Full bridge type in 4-wire, Full bridge type in 6-wire (supports remote sensing)
Resistance Resistor, Potentiometer
Bridge resistance 50 Ω to 10 kΩ  * Available excitation power varies by selection of element.
Built-in element of the bridge 120 or 350 Ω for the quarter- and half-bridge
* Available excitation power varies by selection of element.
Excitation power Voltage mode 1, 2, 2.5, 5, 10 V DC
* Excitation voltage 5 and 10 V is available when bridge resistance is the 350 Ω or higher.
Current mode Constant current: 0.1 to 20 mA (supported voltage is up to 10 V.)
Zero Adjust for Strain gauge Method Fully automatic (via push button or setting the condition menu)
Max. Range ±10,000με (με:10-6 Strain)
Remote sensing 3- or 4-wire in quarter bridge, 4- or 5-wire in half bridge, 6-wire full bridge
Shunt Calibration Approx. 60 kΩ (120 Ω gauge), Approx. 175 kΩ (350 Ω gauge)
Maximum input voltage Between (+) / (-) terminal 10 V,
Common-mode voltage: 10 Vrms AC
Between channels ((-) terminals) 10 Vp-p
Between channel / GND 60 Vp-p
Max. voltage (withstand) Between channels 1000 Vp-p (1 minute)
Between channel / GND 1000 Vp-p (1 minute)
Isolation Between channel / GND Min. 100 MΩ (at 500 V DC)
Common-mode rejection ratio Min. 80 dB (50/60 Hz, Signal source impedance: Max. 300 Ω)
Frequency response DC to 20 kHz
Filter Low pass Off, Line (1.5 Hz), 3, 6, 10, 30, 50, 60, 100, 300, 500 Hz, 1k, 3k, 5k, 10k Hz (in -30dB/oct)
Anti-aliasing Off, On
Support TEDS Standard IEEE 1451.4 Class2 (Temperate No.33)
Support Reading information from the sensor and setting it to module
External dimensions (W x D x H) Approx. 49 x 136 x 160mm (Excluding Protection)
Weight Approx. 840 g
Subject to the conditions:
● Room temperature is 23ºC ±5ºC. ● When 30 minutes or more have elapsed
after power was turned on.
● Filter is set to 10 or Line. ● Sampling rate is set to 1 second.
● GND terminal is connected to ground.

 

● Remote sensing is not available when a NDIS connector is used. ● When a bridge box is used,
the connection needs to be 4-wire or 6-wire full bridge. When connecting with a Half bridge
(Opposite side), an additional bridge box is required. ● Bridge excitation: Constant current
drives a strain gauge type sensor or a 4-wire full bridge. ● The shunt calibration is available
only when the connection is using a 3-wire, 4-wire quarter bridge, 5-wire full bridge, or 6-wire
full bridge.