PARTAC Tactical Particulate Profiler

APPLICATIONS:
  • Ambient Monitoring
  • Military & Homeland Security
  • Emergency Response

Description

Specifications:

Operating Principle Counts individual airborne particles utilizing scattered laser light.
Performance
Particle Size Range 8 Channels, programmable in the range 0.3~m -10~m
Concentration Range 0- 9,000,000 particles per cubic foot (317,700 particles/L)
Accuracy ±10%, to calibration aerosol
Sensitivity 0.3 um @ 2 to 1 peak to valley, 2 to 1 SIN
Flow Rate 1.00 LPM
Sample Time 1 second to 1 hour
Calibration Standards Factory calibrated in accordance with ASTM F328 and ASTM F649
Electrical
Light Source Laser Diode, 35 mW, 780 nm
Power 6-15V DC ( 12V DC with flow system )

Average: 140mA (400mA with flow system)

Max: 190mA (450mA with flow system)

Certifications Meets or exceeds CE, ISO, ASTM and JIS international certifications.
Communication
Standards RS-232C full duplex or RS-485 selectable full or half duplex
Protocol Serial (comma delimited ASCII) or MODBUS RTU
Settings 9600 Baud rate, 8 Data Bits, No Parity, 1 Stop Bit, No Flow Control

Built-in MODBUS registers for laser life time, laser fault, pump life

time, pump fault, & filter fault detection

Physical
Size Length: 10.5″ (26.7 cm); Diameter: 4.00″ (10.2 cm)
Weight 4.0 Ib (1.8 kg)
Mounting Met One’s “QuickMount” to 3/4” IPS vertical pipe stub
Environmental
Operating Temperature -10° to +60° C
Storage Temperature -40° to +80° C