Description
| Property | Value |
|---|---|
| Operating range | P7: 0,5 kV – 7 kV
P30: 3 kV – 30 kV P35: 6 kV – 35 kV |
| Coating thickness | P7: 0,03 mm – 1,7 mm
P7: 1 mils – 67 mils P35: 1,4 mm – 11,3 mm P35: 55 mils – 444 mils |
| Voltage | DC |
| Test voltage indication | LC-Display, 3-digit |
| Accuracy of voltage setting | ± 0.1 kV + 3% of reading |
| Dimensions and weight | High voltage probe: 274 mm x 63 mm (l x dia)/550 g 10.8“ x 2.48“/1.2 lbs/19.4 oz
Control unit: 225 mm x 150 mm x 85 mm (L x W x H)/1400 g / 8.87“ x 5.9“ x 3.35“ (L x W x H)/3 Ibs/49 oz. |
| Alarm signal | 90 dB, 0.1 s/Pore, continuous tone in case of short-circuit |
| Signal output | potential free, Umax.: 100 V, Imax.: 0.4 A |
| Storage battery | 4 C-cells, IEC LR 14, 3.5 Ah, NiMH, replaceable |
| Storage battery life at maximum voltage | P7: ~ 20 h
P30: ~ 10 h P35: ~ 10 h |
| Storage battery charging time | ~ 4h quick charge |
| Mains Voltage | 115 V – 230 V 50 Hz – 60 Hzautomatic switch |
| Operating temperature | 0 °C – 50 °C 32 °F – 122 °F |
| Humidity | avoid dew on the surface (refer to DIN 55 670) |
| Standards | DIN 55 670, DIN 50191 (VDE 0104), DIN EN 61010/Part 1 (VDE 0411/Part 1) |